L-LAS-LT-AL Series
Laser distance line sensors (Advanced Line)General technical data ....
GENERAL TECHNICAL DATA:
Software | Windows® PC software: L-LAS-LT-Scope |
Voltage supply | +24VDC (± 10%), reversed polarity protected, overload protected |
Current consumption | <200mA |
Max. switching current | 100mA, short-circuit proof |
Light source | L-LAS-LT-20-AL L-LAS-LT-38-AL L-LAS-LT-50-AL L-LAS-LT-120-AL ►Halbleiterlaser, 670 nm, <0,39mW opt. Leistung, DC-Betrieb, Laserklasse 1 gem. DIN EN 60825-1:2015-07 (IEC 60825-1:2014) L-LAS-LT-165-AL L-LAS-LT-250-AL ►Halbleiterlaser, 670 nm, <1mW opt. Leistung, DC-Betrieb, Laserklasse 2 gem. DIN EN 60825-1:2015-07 (IEC 60825-1:2014) |
Line detector | CCD line detector with 512 pixels, 2048 subpixels |
Reference distance | L-LAS-LT-20-AL: 19.5mm L-LAS-LT-38-AL: 40mm L-LAS-LT-50-AL: 50mm L-LAS-LT-120-AL: 120mm L-LAS-LT-165-AL: 160mm L-LAS-LT-250-AL: 250mm |
Measuring range | depends on sensor type |
Resolution/reproducibility | depends on sensor type |
Linearity | depends on sensor type |
Laser line geometry | depends on sensor type |
Optical filter | Interference filter, red light filter |
Analog output (1x) | ANA: Voltage 0...+10V or current 4...20mA |
Digital outputs (3x) | OUT0: (-) Measuring value < lower tolerance threshold OUT1: (+) Measuring value > upper tolerance threshold OUT2: (ok) Measuring value within tolerance window pnp bright-switching/npn darks-switching or pnp dark-switching/npn bright-switching Output polarity can be adjusted under Windows® on PC |
Digital inputs (2x) | IN0: External trigger IN1: Teach/Reset (double function) Input voltage +Ub/0V, with protective circuit |
Sensititivy | adjustable under Windows® on PC |
Laser power correction | adjustable under Windows® on PC |
Temperature stability | 0.01% of measuring range/°C |
LED display | 2x two-color LED red/green and 1x two-color LED yellow/green for tolerance band monitoring: red (+): Measuring value > upper tolerance threshold red (-): Measuring value < lower tolerance threshold green (ok): Measuring value within tolerance window 1x two-color LED yellow/green (multifunctional): yellow: power indicator |
Scan frequency | max. 2kHz (at video speed = NORMAL) or max. 3.8kHz (at video speed = FAST) |
Interface | RS232, parameterizable under Windows® |
Housing dimensions | depends on sensor type |
Housing material | Aluminum, anodized in black |
Type of connector | - to PLC: 8-pole fem. connector Binder 712 - to PC: 4-pole fem. connector Binder 707 |
Connecting cables | - to PLC: cab-las8/SPS - to PC: cab-las4/PC, cab-4/USB or cab-4/ETH |
Operating temp. range | -10°C ... +50°C |
Storage temp. range | -20°C ... +85°C |
Enclosure rating | Electronics: IP54, optics: IP67 |
EMC test acc. to | DIN EN 60947-5-2 |